PSTatistic Subsystem 9
Keysight N8262A Programming Guide 333
PSTatistic[1]|2:CCDF:POWer? <numeric_value>
This command is used to return the power level at the specified probability.
Syntax
Parameters
Example
NOTE
This command is only applicable when P-Series sensors are present and the
following conditions are met:
Free run acquisition mode is selected
NORMal or DOUBle measurement speed setting is chosen
PST
2
1
:CCDF
?
:POW
numeric_value
space
Item Description/Default Range of Values
numeric_value The probability at the queried power.
Maximum value: 0 %
Minimum value: 100 %
0.0 to 100
PST1:CCDF:POW? 30 This command queries the power level at
probability of 30 %.