TRIGger Subsystem 14
Keysight N8262A Programming Guide 573
INITiate:CONTinuous:ALL <boolean>
Sets all trigger sequences to be continuously initiated.
If INITiate:CONTinuous:ALL is set to:
ON, trigger sequences are set to be continuously initiated
OFF, trigger sequences are not set to be continuously initiated
Syntax
Example
Reset Condition
On reset (*RST), this command is set to OFF.
On preset (SYSTem:PRESet) and instrument power-up, when entering local mode,
if TRIGger[:SEQuence[1]|2]:SOURce is set to INT[[1]|2] or EXT,
INITiate:CONTinuous is not changed. For other trigger sources,
INITiate:CONTinuous is set to ON.
INIT:CONT:ALL ON This command sets all trigger sequences to
be continuously initiated.
INIT
:CONT
0|OFF
1|ON
Space
?
:ALL