TRIGger Subsystem 14
Keysight N8262A Programming Guide 585
Query
TRIGger:SOURce?
The query returns the current trigger source, either IMM, BUS or HOLD.
Query Example
Error Messages
– If the master is changed to IMM, BUS or HOLD, error –221 “Settings Conflict”
occurs. In such situations the slave’s TRIG:SOUR must be changed so that it is
no longer a slave.
– If the source is changed to INT1, INT2 or EXT and SENS:SPEED has a value of
200, error –221 “Settings Conflict” occurs.
– If the source is changed to INT1, INT2 or EXT and SENS:DET:FUNC is set to
AVERage, error –221 “Settings Conflict” occurs.
TRIG:SOUR? This command queries Channel A’s trigger
source.