12 SYSTem Subsystem
500 Keysight N8262A Programming Guide
TRIG[:SEQ]:LEV –15 dBm Power level
TRIG[:SEQ]:SLOP POS Trigger event recognized on the rising
edge of a signal
TRIG[:SEQ]:DEL 0 s Delay between recognition of trigger event
and start of a measurement
TRIG[:SEQ]:HOLD 650 µs Trigger holdoff
Range
[a]
[SENS[1]]|SENS2:POW:AC:RANG:AUTO OFF Auto range off
[SENS[1]]|SENS2:POW:AC:RANG UPPER Range set to upper
Step detection
[SENSe[1]]|SENS2:AVER:SDET 1 Step detection enabled
Trace setup
SENSe[1]|2:TRAC:LIM:UPP +20 dBm Maximum power
SENSe[1]|2:TRAC:LIM:LOW –35 dBm Minimum power
[SENS[1]]|SENS2:TRAC:OFFS:TIME
<numeric_value>
–50 µs Delay between delayed trigger point and
the start of the trace
[SENS[1]]|SENS2:TRAC:TIME
<numeric_value>
3.8 ms Length of the trace
[a] The Range setting in Table 12-17 is only applicable for E-Series power sensor.
Table 12-17 BLUetooth: Power meter presets (continued)
Command Setting Comments