STATus Subsystem 11
Keysight N8262A Programming Guide 431
STATus:OPERation:CALibrating[:SUMMary]
The operation status calibrating summary register set contains information on the
calibrating status of the power meter.
The following bits in these registers are used by the power meter:
These bits are set at the beginning of zeroing (CALibration:ZERO:AUTO ONCE) and
at the beginning of calibration (CALibration:AUTO ONCE). Also for the compound
command/query CALibration[:ALL]?, this bit is set at the beginning of the
calibration sequence.
These bits are cleared at the end of zeroing or calibration.
Syntax
Bit Number Decimal Weight Definition
0-Not used
1 2 Channel A CALibrating Status
2 4 Channel B CALibrating Status
3-15 - Not used
STAT
:OPER
:CAL
:SUMM