MEMory Subsystem 7
Keysight N8262A Programming Guide 305
MEMory:TABLe:FREQuency <numeric_value>{,<numeric_value>}
This command is used to enter frequency data into the current selected table. Any
previous frequency list is cleared before the new frequency list is stored. The
frequencies must be entered in ascending order. Entries in the frequency lists
correspond as shown in Table 7-3 with entries in the calibration/offset factor lists.
For sensor calibration tables only, the first calibration factor entered using the
MEMory:TABLe:GAIN command is used as the reference calibration factor.
Table 7-3 Frequency and Calibration/Offset factor list
Table Power Sensor Table Name
0None
DEFAULT
[a]
1 8481A 8481A
2 8482A, 8482B, 8482H 8482A
3 8483A 8483A
4 8481D 8481D
5 8485A 8485A
6 R8486A R8486A
7 Q8486A Q8486A
8 R8486D R8486D
9 8487A 8487A
[a] Default is a sensor calibration table in which the reference calibration factor and calibration factors are 100%. This sensor calibration table
can be used during the performance testing of the power meter.