OUTPut Subsystem 8
Keysight N8262A Programming Guide 325
OUTPut:TRIGger[:STATe] <boolean>
This command enables/disables the trigger output signal.
Syntax
Example
Reset Condition
On reset, the trigger output signal is disabled.
Query
OUTPut:TRIGger[:STATe]?
The query command enters a 1 or 0 into the output buffer indicating whether or
not the trigger output signal is enabled/disabled.
1 is returned when the trigger output signal is enabled
0 is returned when the trigger output signal is disabled
Query Example
OUTP:TRIG:STAT 1 This command enables the trigger output
signal.
OUTP
:STAT
:TRIG
0|OFF
1|ON
Space
?
OUTP:TRIG:STAT? This command queries the status of the
trigger output signal.