Power Meter Remote Operation 1
Keysight N8262A Programming Guide 105
Using the Operation Complete Commands
The *OPC? and *OPC commands allow you to maintain synchronization between
the computer and the power meter. The *OPC? query command places an ASCII
character 1 into the power meter’s output queue when all pending power meter
commands are complete. If your program reads this response before continuing
program execution, you can ensure synchronization between one or more
instruments and the computer.
The *OPC command sets bit 0 (Operation Complete) in the Standard Event Status
Register when all pending power meter operations are complete.
Procedure
– Send a device clear message to clear the power meter’s output buffer.
– Clear the event registers with the *CLS (clear status) command.
– Enable operation complete using the *ESE 1 command (standard event
register).
– Send the *OPC? (operation complete query) command and enter the result to
assure synchronization.
– Send your programming command string, and place the *OPC (operation
complete) command as the last command.
– Send the *STB? (status byte query) command to poll the register. This
command does not clear the status byte summary register.
Examples
This example program uses the *OPC? command to determine when the power
meter has finished calibrating.
CAL:AUTO ONCE
*OPC?
MEAS:POW:AC?