EasyManuals Logo

Keysight Technologies P Series User Manual

Keysight Technologies P Series
673 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #572 background imageLoading...
Page #572 background image
14 TRIGger Subsystem
572 Keysight N8262A Programming Guide
INITiate[1]|2[:IMMediate]
This command sets the power meter in the wait for trigger state. When a trigger is
received, the measurement is taken and the result placed in the power meter
memory. If TRIGger:SOURce is set to IMMediate the measurement begins as soon
as INITiate:IMMediate is executed.
Use FETCh? to transfer a measurement from memory to the output buffer. Refer to
“FETCh[1]|2|3|4 Queries” on page 145 for further details.
Syntax
Example
Error Messages
If the power meter is not in the idle state or INITiate:CONTinuous is ON, error
–213, “INIT ignored” occurs.
NOTE
This command performs the same function as
INITiate:[IMMediate]:SEQuence[1]|2.
INIT2:IMM This command places Channel B in the wait
for trigger state.
INIT
:IMM
1
2

Table of Contents