EasyManuals Logo

Keysight Technologies P Series User Manual

Keysight Technologies P Series
673 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #615 background imageLoading...
Page #615 background image
SERVice Subsystem 16
Keysight N8262A Programming Guide 615
SERVice:BIST:CALibrator <boolean>
This command enables/disables the calibrator self-test during power-up. It can be
used to disable the self-test if it incorrectly indicates a failure. If a load, for example,
a sensor, is connected to the calibrator port this could cause the self-test to fail.
Also, if it fails the self-test, a Pop-up is displayed for 5 seconds, stating -If Ref
Calibrator test fails disconnect any load attached to it and re-try test.
Syntax
Example
Query
SERVice:BIST:CALibrator?
The query enters a 1 or 0 into the output buffer indicating the status of the
self-test.
– 1 is returned when the self-test is enabled
– 0 is returned when the self-test is disabled
Query Example
SERV:BIST:CAL OFF This command disables the calibrator
self-test during power-up.
:BIST
SERV 0|OFF
1|ON
Space
?
:CAL
ONCE
SERV:BIST:CAL? This command queries whether the self-test
is enabled or disabled.

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals