EasyManuals Logo

Keysight Technologies P Series User Manual

Keysight Technologies P Series
673 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #252 background imageLoading...
Page #252 background image
5 DISPlay Subsystem
252 Keysight N8262A Programming Guide
DISPlay[:WINDow[1]|2]:ANALog:UPPer <numeric_value>
This command sets the analog meter upper scale limit.
The units used are dependent on the current setting of UNIT:POWer and
CALCulate:RELative:STATe as shown in Table 5-2.
NOTE
This command has the same purpose as DISPlay[:WINDow[1]|2]:METer:UPPer
<numeric_value>.
Table 5-2 Measurement units
Measurement Mode Measurement Type CALC:REL:STAT OFF CALC:REL:STAT ON
Linear Log Linear Log
Single Channel
Avg, Pk Watt dBm % dB
Pk–Avg % dB % dB
Ratio
Avg, Pk, Pk–Avg % dB % dB
Difference
Avg, Pk Watt dBm % dB
Pk–Avg % dB % dB

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals