SYSTem Subsystem 12
Keysight N8262A Programming Guide 525
The following table shows the power meter presets when <character_data> is
set to NADC. Commands not listed are preset according to their DEFault values (for
further information refer to Table 12-1):
Table 12-41 NADC: Power meter presets
Command Setting Comments
Frequency
[SENS[1]]|SENS2:FREQ[:CW|:FIX] +800.000 MHz Frequency setting
Sensor measurement mode
[SENS[1]]|SENS2:DET:FUNC NORM Measurement mode
Sensor video bandwidth setup
[SENS[1]]|SENS2:BAND|BWID:VID N1921/2A: OFF Sensor video bandwidth
Gate Setup
[SENS[1]]|SENS2:SWE[1]|2|3|4:OFF:TIME
Gate 1: 123.5
µs
Gate 2: 20.123 ms
Gates 3 - 4: 0
Delay between trigger point and time
gated period.
[SENS[1]]|SENS2:SWE[1]|2|3|4:TIME Gate 1: 6.46 ms
Gate 2: 6.46 ms
Gates 3 - 4: 0
Length of time gated period for time gated
measurements.
Trigger setup
TRIG[:SEQ[1]|2]:SOUR INT1 Trigger source set up and acquisition
mode continuous triggering
INIT:CONT ON
TRIG[:SEQ]:LEV:AUTO OFF Disable automatic setting of the trigger
level
TRIG[:SEQ]:LEV -15 dBm Power level
TRIG[:SEQ]:SLOP POS Trigger event recognized on the rising
edge of a signal
TRIG[:SEQ]:DEL 0 s Delay between recognition of trigger event
and start of a measurement
TRIG[:SEQ]:HOLD 30 ms Trigger holdoff
Range