7 MEMory Subsystem
306 Keysight N8262A Programming Guide
For sensosr calibration tables (8480 Series sensors only), the number of frequency
points must be one less than the number of calibration factor points. This is
verified when the sensor calibration table is selected using
SENSe:CORRection:CSET:SELect <string>.
Ensure that the frequency points you use cover the frequency range of the signals
that you want to measure. If you measure a signal with a frequency outside the
frequency range defined in the table, then the power meter uses the highest or
lowest point in the table to calculate the calibration factor/offset.
Depending on available memory, the power meter is capable of storing 20 sensor
calibration tables and 10 frequency dependent offset tables, each containing 80
points.
Syntax
Parameters
MEM
:TABL
:FREQ
numeric_value
Space
?
,
Item Description/Default Range of Values
numeric_value A numeric value for the frequency. The default units
are Hz.
1 kHz to 1000.0 GHz
[a],[b]
[a] The following measurement units can be used:
Hz
kHz (10
3
)
MHz (10
6
)
GHz (10
9
)
[b] All frequencies are truncated to a multiple of 1 kHz.

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals