SENSe Subsystem 10
Keysight N8262A Programming Guide 375
Syntax
Parameters
Example
SENS
1
:DCYC
CORR
SENS2
:
:INP
:MAGN
Space
?
numeric_value
DEF
MIN
MAX
MIN
MAX
Space
:GAIN3
Item Description/Default Range of Values
numeric_value A numeric value for the duty cycle.
– DEF: the default value is 1 %
– MIN: 0.001 %
– MAX: 99.999 %
The units are PCT, and are optional.
0.001 to 99.999 PCT
DEF
MIN
MAX
CORR:DCYC 90PCT This command sets a duty cycle of 90 % for
Channel A.