CALibration Subsystem 4
Keysight N8262A Programming Guide 231
Syntax
Example
Reset Condition
On reset, automatic calibration is disabled.
Query
CALibration[1]|2:AUTO?
The query always returns a value of 0.
Error Messages
If this command is set to ON and an 8480 Series or E-Series power sensor is
connected the error –241, “Hardware missing” occurs.
If the calibration was not carried out successfully the error –231, “Data
Questionable; CAL ERROR” occurs. The error message will specify which
channel failed calibration.
If there is no sensor connected, the error –241, “Hardware Missing” occurs.
CAL1:AUTO ONCE This command causes the power meter to
perform a calibration on Channel A.
:AUTO
0|OFF
1|ON
Space
?
CAL
1
2
ONCE