SENSe Subsystem 10
Keysight N8262A Programming Guide 359
Query
AVERage2:COUNt?
The query returns the current setting of the video filter length. The format of the
response is <NR1>.
Query Example
Error Messages
– If the command is used when a sensor other than a P-Series power sensor is
connected, error –241, “Hardware missing” occurs
AVER2:COUN? This command queries the video filter length
for Channel A.