10 SENSe Subsystem
360 Keysight N8262A Programming Guide
[SENSe[1]]|SENSe2:AVERage2[:STATe] <boolean>
This command is used to enable and disable video averaging for the P-Series
sensor.
Syntax
Example
Reset Condition
On reset, averaging is enabled.
Query
[SENSe[1]]|SENSe2:AVERage2[:STATe]?
The query enters a 1 or 0 into the output buffer indicating the status of averaging.
– 1 is returned when averaging is enabled.
– 0 is returned when averaging is disabled.
AVER2 1 This command enables video averaging on
Channel A.
SENS
1
:STAT
AVER2 0|OFF
1|ON
Space
?
SENS2
: