TRACe Subsystem 13
Keysight N8262A Programming Guide 555
TRACe[1]|2:MEASurement:TRANsition[1]|...|10:NEGative:OCCurrence?
This command returns the position, relative to the trigger instant, of the n
th
occurrence of a negative transition found on a trace.
Syntax
Reset Condition
On reset, this parameter is not affected.
Example
Error Messages
– If a P-Series sensor is not connected, error –241, “Hardware missing” occurs.
– If a P-Series sensor is connected and Free Run trigger acquisition is selected,
error –221, “Settings conflict” occurs.
TRAC
:MEAS
1
2
:TRAN
:NEG
:OCC
?
1
2|...|10
TRAC2:MEAS:TRAN7:NEG:OCC? This command returns the position, relative to
the trigger instant, of the 7th occurrence of a
negative transition found on trace 2.
If you attempt to measure a pulse out of the range of the capture, for example,
measure the 5th pulse and there are only 4 pulses displayed, the power meter
returns #0#0#0 as the result.