14 TRIGger Subsystem
582 Keysight N8262A Programming Guide
TRIGger[1]|2[:IMMediate]
This command causes a trigger to occur immediately, provided the specified
channel is in the wait for trigger state. When this command is executed, the
measurement result is stored in the power meter’s memory. Use FETCh? to place
the measurement result in the output buffer.
TRIGger[1]|2:DELay:AUTO is ignored if TRIGger[1]|2[:IMMediate] is used to
trigger the power meter.
Syntax
Example
Error Messages
If the power meter is not in the wait for trigger state, then TRIGger:IMMediate
causes error –211, “Trigger ignored”.
This command performs the same function as INITiate[1]|2:[IMMediate].
TRIG This command causes a Channel A trigger to
occur immediately.