EasyManuals Logo

Keysight Technologies P Series User Manual

Keysight Technologies P Series
673 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #638 background imageLoading...
Page #638 background image
16 SERVice Subsystem
638 Keysight N8262A Programming Guide
SERVice:SENSor[1]|2:POWer:PEAK:MAXimum?
This query returns the maximum peak power that can be measured by the
currently connected sensor. It is applicable to E-Series sensors only. Maximum
peak power information is stored in the sensor’s EEPROM.
Syntax
Example
Error Messages
If no sensor is connected, error –241, “Hardware missing” occurs.
If a a sensor other than an E-Series power sensor is connected, error –241
“Hardware missing” occurs.
If the E-Series sensor currently connected does not contain the necessary
information in EEPROM, error –241 “Hardware missing” occurs.
SERV:SENS2:POW:PEAK:MAX? This query returns the maximum peak power
that can be measured by the E-Series sensor
currently connected to Channel B.
:POW
?
:PEAK
:MAX
SERV
1
2
:SENS

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals