2 MEASurement Commands
132 Keysight N8262A Programming Guide
CONFigure [1] |2|3|4 Commands
The CONFigure commands are used on the specified window/measurement to set:
– The expected power level being measured.
– The resolution of the window/measurement.
– The channel(s) on which the measurement is to be made.
The CONFigure commands do not make the power measurement after setting the
configuration. Use READ?, or alternatively use INITiate followed by a FETCh? to
make the measurement.
The CONFigure command also applies the following defaults to the channel(s)
which are in the specified window (the channel(s) in the window are specified in
the <source list> parameter):
Default Settings Description
INITiate:CONTinuous OFF Sets the power meter to make one trigger cycle when INITiate
is sent.
TRIGger:SOURce IMMediate When TRIG:SOUR is set to BUS or HOLD, sets the power meter
to make the measurement immediately a trigger is received.
TRIGger:DELay:AUTO ON Enables automatic delay before making the measurement.
SENSE:AVERage:COUNt:AUTO ON Enables automatic filter length selection.
SENSE:AVERage:STATe ON Enables averaging.

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals