14 TRIGger Subsystem
570 Keysight N8262A Programming Guide
INITiate[1]|2:CONTinuous <boolean>
This command sets the power meter for either a single trigger cycle or continuous
trigger cycles. A trigger cycle means that the power meter exits the wait for
trigger state and starts a measurement.
When entering local mode, if TRIGger[:SEQuence[1]|2]:SOURce is set to
INT[[1]|2] or EXT, INITiate:CONTinuous is not changed. For other trigger
sources, INITiate:CONTinuous is set to ON.
If INITiate:CONTinuous is set to:
– OFF, the trigger system remains in the idle state until it is set to ON, or
INITiate:IMMediate is received. Once this trigger cycle is complete the
trigger system returns to the idle state.
– ON, the trigger system is initiated and exits the idle state. On completion of
each trigger cycle, the trigger system immediately commences another trigger
cycle without entering the idle state.
Syntax
NOTE
This command performs the same function as
INITiate:CONTinuous:SEQuence[1]|2 <boolean>.
INIT
1
2
:CONT 0|OFF
1|ON
Space
?

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals