SERVice Subsystem 16
Keysight N8262A Programming Guide 627
SERVice:CALibrator:ADJ:COUR <numeric_value>
This command adjust the 1 mW calibrator output in coarse scale.
Syntax
Parameters
Query
:CAL
:ADJ
SERV
:COUR
Space
numeric_value
DEF
?
Item Description/Default Range of Values
Numeric_value Adjust the 1 mW Power Reference Level Increment Coarse by 1. 0 to 1023 (Unsigned Int 16)
SERV:CAL:ADJ:COUR? The query returns the Reference Calibrator
power level in unsigned Int 16.