16 SERVice Subsystem
624 Keysight N8262A Programming Guide
SERVice:BIST:TBASe:STATe <boolean>
This command sends a 10 MHz time base signal to the front panel trig out for
testing purposes.
If the command is set to:
– ON, the 10 MHz time base signal is sent to the front panel trigger out
connector.
– OFF, the 10 MHz time base signal is disabled.
Syntax
Example
NOTE
This command overrides the OUTPut:TRIGger[:STATe] command.
For example, if OUTPut:TRIGger[:STATe] is ON and the command
SERV:BIST:TBAS ON is sent, this command overrides the Trigger state and sets it
to OFF. However, the 10 MHz remains out the Trig out port.
If the SERV:BIST:TBAS ON has been sent, the 10 MHz is on and the
OUTPut:TRIGger[:STATe] is then toggled to ON, the channel trigger is now
routed to the Trig out overriding the service command turning the 10 MHz to off.
SERV:BIST:TBAS:STAT OFF This command disables the signal.
:BIST
SERV
0|OFF
1|ON
Space
?
:TBAS
:STAT

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals