1 Power Meter Remote Operation
106 Keysight N8262A Programming Guide
Saving and Recalling Power Meter Configurations
To reduce repeated programming, up to ten power meter configurations can be
stored in the power meter’s non-volatile memory. The error list, remote addresses,
sensor calibration table data, zeroing and calibration information are not stored.
How to Save and Recall a Configuration
Power meter configurations are saved and recalled with the following commands:
*SAV <NRf>
*RCL <NRf>
The range of values for <NRf> in the above commands is 1 to 10.
Example Program
10 ASSIGN @POWER TO 713
20 !Configure the power meter
30 OUTPUT @POWER;“UNIT:POW W”
40 OUTPUT @POWER;“SENS:CORR:LOSS2 -10”
50 OUTPUT @POWER;“SENS:CORR:LOSS2:STAT ON”
60 !Save the configuration
70 OUTPUT @POWER;“*SAV 5”
80 PRINT “Configuration Saved”
90 !Now reset the power meter
100 OUTPUT @POWER;“*RST”
110 ! Recall the configuration
120 OUTPUT @POWER;”*RCL 5”
130 PRINT “Configuration Recalled”
140 PRINT “Save and Recall complete”
150 END

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals