7 MEMory Subsystem
288 Keysight N8262A Programming Guide
MEMory:CATalog:TABLe?
This command is used to list the stored sensor calibration (8480 Series sensors
only) and frequency dependent offset tables.
The power meter returns the data in the form of two numeric parameters and as
many strings as there are stored tables.
<numeric_value>,<numeric_value>{,<string>}
– The first numeric parameter indicates the amount of memory, in bytes, used
for the storage of tables.
– The second parameter indicates the memory, in bytes, available for the
storage of tables.
– Each string parameter returned indicates the name, type and size of a stored
table:
– <string>,<type>,<size>
– <string> indicates the name of the table.
– <type> indicates TABL for a table.
– <size> indicates the size of the table in bytes.
For example, a sample of a response may look like:
1178,10040,"DEFAULT,TABL,14","8481A,TABL,116",
"8482A,TABL,74","8483A,TABL,62"...........
The power meter is shipped with a set of predefined sensor calibration tables. The
data in these sensor calibration tables is based on statistical averages for a range
of Keysight Technologies power sensors. These tables can be edited. The
predefined data is listed in your User’s Guide. These power sensors and table
numbers are listed in Table 7-2.