SENSe Subsystem 10
Keysight N8262A Programming Guide 397
Example
Reset Condition
On reset, the value is set to 0 seconds.
Query
SENSe[1]]|SENSe2:SWEep[1]|2|3|4:OFFSet:TIME?
The query returns the current delay between the trigger point and the start of the
time-gated period.
Query Example
Error Messages
If the command is used when a sensor other than a P-Series power sensor is
connected, error –241, “Hardware missing” occurs
SENS2:SWE3:OFFS:TIME 0.001 This command sets the delay to 0.001
seconds.
SENS2:SWE2:OFFS:TIME? The query returns the current delay between
the trigger point and the start of the
time-gated period for Channel B and gate 2.