CALCulate Subsystem 3
Keysight N8262A Programming Guide 189
CALCulate[1]|2|3|4:FEED[1]|2 <string>
This command sets the input measurement mode to be fed to the specified input
on the CALC block. It is applied to the measurement after the CALC:MATH:EXPR
command has been used to specify which channel the feed is taken from.
Measurement modes are coupled for combination measurements (for example,
ratio measurements). For example, if one feed is changed to PTAV, the other is
automatically changed to PTAV.
Under certain circumstances the measurement mode is changed by the
CALC:MATH:EXPR command. Refer to “CALCulate[1]|2|3|4:MATH[:EXPRession]
<string>” on page 212 for further information.
Syntax
CALC
Space
?
string
:FEED
1
2
1
2
3
4