TRACe Subsystem 13
Keysight N8262A Programming Guide 543
Example
Reset condition
On reset, the reference level will become 50 %, which is the default value (DEF).
Query
TRACe[1]|2:DEFine:DURation:REFerence?
The query returns the numeric value of the reference level used in the pulse
duration calculation.
Query Example
TRAC1:DEF:DUR:REF 25 This command sets trace 1 pulse duration
measurements to look for the 25 % reference
levels.
TRAC1:DEF:DUR:REF DEF This command sets trace 1 pulse duration
measurements to look for the 50 % reference
levels.
TRAC1:DEF:DUR:REF? This command queries the value of the
reference level used in pulse duration
measurement for trace 1.