2 MEASurement Commands
164 Keysight N8262A Programming Guide
Example
Error Messages
INITiate:CONTinuous must be set to OFF on both channels, otherwise error
–213, “INIT ignored” occurs.
–If TRIGger:SOURce is set to BUS or HOLD on either channel, error –214, “Trigger
deadlock” occurs.
If the resolution parameter is not the same as the current resolution setting on
the specified window, error –221, “Settings conflict” occurs.
Item Description/Default Range of Values
expected_value
(for the expected power
level)
The power meter ignores the numeric value entered in this
parameter. Any value entered is treated like DEF.
Sensor dependent
DEF
[a]
resolution A numeric value for the resolution. If it is unspecified the current
resolution setting is used. If a value is entered it should correspond
to the current resolution setting otherwise an error occurs.
1 to 4
[b]
1.0, 0.1, 0.01, 0.001
DEF
[a]
source list This channel list specifies the channels used to calculate the
difference.
If unspecified and the current window setup is a difference
measurement then this difference setup is used, otherwise it
defaults to Channel A-B.
(@1),(@2)
(@2),(@1)
(@1),(@1)
(@2),(@2)
[a] The mnemonic DEF means DEFault. This is not equivalent to the DEFault parameter used in the command sub-systems. The parameters must
be entered in the specified order. If parameters are omitted, they default from the right. The parameter DEFault is used as a place holder.
Specifying DEF leaves the parameter value unchanged.
[b] When the measurement result is linear this parameter represents the number of significant digits. When the measurement result is
logarithmic 1 to 4 represents of 1, 0.1, 0.01 and 0.001 respectively.
READ2:DIFF? This command queries difference
measurement on the lower window/upper
measurement.

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals