5 DISPlay Subsystem
262 Keysight N8262A Programming Guide
DISPlay[:WINDow[1]|2]:METer:UPPer <numeric_value>
This command sets the analog meter upper scale limit.
The units used are dependent on the current setting of UNIT:POWer and
CALCulate:RELative:STATe as shown in Table 5-4.
NOTE
This command has the same purpose as DISPlay[:WINDow[1]|2]:ANALog:UPPer
<numeric_value>.
Table 5-4 Measurement units
Measurement Mode Measurement Type CALC:REL:STAT OFF CALC:REL:STAT ON
Linear Log Linear Log
Single Channel Avg, Pk Watt dBm % dB
Pk–Avg % dB % dB
Ratio Avg, Pk, Pk–Avg % dB % dB
Difference Avg, Pk Watt dBm % dB
Pk–Avg % dB % dB