CALCulate Subsystem 3
Keysight N8262A Programming Guide 193
CALCulate[1]|2|3|4:GAIN[:MAGNitude] <numeric_value>
This command is used to enter a value for the display offset on the specified
window/measurement. The display offset is applied to the measurement signal
after any math calculation.
Entering a value using this command automatically turns the
CALCulate[1]|2|3|4:GAIN:STATe command to ON.
Syntax
Parameters
CALC
:MAGN
:GAIN
Space
?
numeric_value
DEF
MIN
MAX
MIN
MAX
Space
1
2
3
4
Item Description/Default Range of Values
numeric_value A numeric value for the display offset:
– DEF: the default value is 0 dB
– MIN: –100.000 dB
– MAX: +100.000 dB
–100.000 to +100.000 dB
DEF
MIN
MAX