Power Meter Remote Operation 1
Keysight N8262A Programming Guide 85
Auto Averaging Mode
To enable and disable auto filter mode, use the following command:
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO <boolean>
When the auto filter mode is enabled, the power meter automatically sets the
number of readings averaged together to satisfy the filtering requirements for
most power measurements. The number of readings averaged together depends
on the resolution and the power level currently being measured. Figure 1-2 lists
the number of readings averaged for each range and resolution when the power
meter is in auto filter mode.
Figure 1-2 Typical averaged readings on 8480 Series sensors
NOTE
Figure 1-2 applies to 8480 Series sensors only.
10 dB
10 dB
Minimum Sensor Power
Maximum Sensor Power
Power Sensor
Dynamic Range
10 dB
10 dB
1234
8 8 128 128
Resolution Setting
Number of Averages
1 1 16 256
11232
11116
1118

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals