12 SYSTem Subsystem
488 Keysight N8262A Programming Guide
Trigger setup
TRIG[:SEQ[1]|2]:SOUR INT1 Trigger source set up and acquisition
mode continuous triggering
INIT:CONT ON
TRIG[:SEQ]:LEV:AUTO ON Enable automatic setting of the trigger
level
TRIG[:SEQ]:LEV AUTO Automatic Power level
TRIG[:SEQ]:SLOP POS Trigger event recognized on the rising
edge of a signal
TRIG[:SEQ]:DEL 0 s Delay between recognition of trigger event
and start of a measurement
TRIG[:SEQ]:HOLD MIN Trigger holdoff
Range
[a]
[SENS[1]]|SENS2:POW:AC:RANG:AUTO OFF Auto range off
[SENS[1]]|SENS2:POW:AC:RANG UPPER Range set to upper
Step detection
[SENSe[1]]|SENS2:AVER:SDET 0 Step detection disabled
[a]
The Range setting in Table 12-8 is only applicable for E-Series power sensor.
Table 12-8 CDMAone: Power meter presets (continued)
Command Setting Comments