17 IEEE 488.2 Command Reference
654 Keysight N8262A Programming Guide
*DDT <arbitrary block program data>|<string program data>
The *DDT (Define Device Trigger) command determines the power meter’s
response to a GET (Group Execute Trigger) message or *TRG common command.
This command effectively turns GET and *TRG into queries, with the measured
power being returned.
Syntax
Parameters
Examples of <arbitrary block program data> parameters are:
#15FETC? and #206FETCh?
string program data
*DDT
Space
arbitrary block program data
?
Type Description Range of Values
arbitrary block program data
The command which is executed on a
GET or *TRG.
#nN<action>
[a],[b]
string program data
“<action>”
[a]
[a] The <action> field of the parameter may contain:
FETC?
FETC1?
FETC2?
*TRG
TRIG1
TRIG2
[b] The first digit after the # indicates the number of following digits. The following digits indicate the length of the data.

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals