SENSe Subsystem 10
Keysight N8262A Programming Guide 417
Example
Reset Condition
On reset, Y Max = 20 dBm; Y Scale = 7 dB.
Query
SENSe[1]|2:TRACe:Y:SCALe:PDIV?
The query returns the current scale setting of Y-axis in numerical value.
Query Example
Error Messages
– If no sensor or wrong sensor is connected to the channel, error –241
"Hardware missing" occurs.
– If channel is not in trigger acquisition mode, error –221 "Settings conflict"
occurs.
– If limits of the values keyed in are exceeded, error –222 "Data out of range;
upper (or lower) limit exceeded; no change" occurs.
SENS:TRAC:Y:SCAL:PDIV 0.002 This command sets the Channel A Y Scale
value to 0.002 step.
SENS2:TRAC:Y:SCAL:PDIV 0.05 This command sets the Channel B Y Scale
value to 0.05 step.
SENS:TRAC:Y:SCAL:PDIV? This command queries the Channel A current
Y-axis scale setting in numerical value.

Table of Contents