12 SYSTem Subsystem
476 Keysight N8262A Programming Guide
Preset Values
DEFault
Table 12-1 shows the power meter presets when <character_data> is set to
DEFault or omitted. Values are shown for all SCPI commands:
Table 12-1 DEFault: Power meter presets
Command Setting Comments
CALC[1]|2|3|4:FEED[1]|2 “POW:AVER” Select average measurement type
CALC[1]|2|3|4:GAIN[:MAGN] 0.000 dB Display offset value
CALC[1]|2|3|4:GAIN:STAT OFF Display offset disabled
CALC[1]|2|3|4:LIM:CLE:AUTO ON Clear limit data at INIT
CALC[1]|2|3|4:LIM:LOW[:DATA] –90 dBm Lower limit
CALC[1]|2|3|4:LIM:STAT OFF Window limits checking disabled
CALC[1]|2|3|4:LIM:UPP[:DATA] +90 dBm
CALC[1]|2|3|4:MATH[:EXPR] Upper - Channel A
Lower - Channel B
Math expression
CALC[1]|2|3|4:REL[:MAGN]:AUTO OFF Reference value disabled
CALC[1]|2|3|4:REL:STAT OFF Relative offset disabled
CAL[1]|2:ECON:STAT OFF TTL zero/calibration inputs disabled
CAL[1]|2:RCAL not affected zero/cal lockout
CAL[1]|2:RCF 100.0 % Reference calibration factor
DISP:CONT not affected Display contrast
DISP:ENAB ON Display enabled
DISP:SCR:FORM WIND Display format set to windowed
DISP[:WIND[1]|2]:ANAL:LOW –70 dBm Lower scale limit
DISP[:WIND[1]|2]:ANAL:UPP 20 dBm Upper scale limit
DISP[:WIND[1]|2]:FORM Upper - digital
Lower - digital
Display format