CALCulate Subsystem 3
Keysight N8262A Programming Guide 205
Parameters
Example
Reset Condition
On reset, both measurements of the upper and lower windows are set to
–90.00 dBm or –90 dB (DEF).
Query
CALCulate[1]|2|3|4:LIMit:LOWer[:DATA]? [MIN|MAX]
The query returns the current setting of the lower limit or the values associated
with MIN and MAX for the specified window.
Query Example
Item Description/Default Range of Values
numeric_value A numeric value for the lower test limit:
DEF: the default is –90.00 dBm or –90 db
MIN: –150 dBm or –180 dB
MAX: +230 dBm or +200 dB
–150 to +230 dBm or
–180 to +200 dB
DEF
MIN
MAX
CALC2:LIM:LOW:DATA 0.1 This command enters a lower limit for the
lower window/upper measurement
depending on the window’s units as follows:
dBm = 0.1 dBm
W = 100 mW
dB = 0.1 dB
% = 0.1 %
CALC2:LIM:LOW:DATA? This command queries the lower limit set for
the lower window upper measurement.