12 SYSTem Subsystem
532 Keysight N8262A Programming Guide
DVB
The following table shows the power meter presets when <character_data> is
set to DVB. Commands not listed are preset according to their DEFault values (for
further information refer to Table 12-1).
Table 12-47 DVB: Power meter presets
Command Setting Comments
Frequency
[SENS[1]]|SENS2:FREQ[:CW|:FIX] +660.000 MHz Frequency setting
Sensor measurement mode
[SENS[1]]|SENS2:DET:FUNC NORM Measurement mode
Sensor video bandwidth setup
[SENS[1]]|SENS2:BAND|BWID:VID N1921/2A: OFF Sensor video bandwidth
Gate Setup
[SENS[1]]|SENS2:SWE[1]|2|3|4:OFF:TIME
Gate 1: 10
µs
Gate 2: 0
Gates 3 - 4: 0
Delay between trigger point and time
gated period.
[SENS[1]]|SENS2:SWE[1]|2|3|4:TIME Gate 1: 15 ms
Gate 1: 90 ms
Gates 2 - 4: 0
Length of time gated period for time gated
measurements.
Trigger setup
TRIG[:SEQ[1]|2]:SOUR INT1 Trigger source set up and acquisition
mode continuous triggering
INIT:CONT ON
TRIG[:SEQ]:LEV:AUTO OFF Disable automatic setting of the trigger
level
TRIG[:SEQ]:LEV -15 dBm Power level
TRIG[:SEQ]:SLOP POS Trigger event recognized on the rising
edge of a signal
TRIG[:SEQ]:DEL 0 s Delay between recognition of trigger event
and start of a measurement
TRIG[:SEQ]:HOLD 20 ms Trigger holdoff
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals