10 SENSe Subsystem
398 Keysight N8262A Programming Guide
[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:TIME <numeric_value>
This command sets the length of the time-gated period (time-gate length) for
time-gated measurements for the P-Series sensors which are set to NORMal
mode.
Syntax
Parameters
Example
SENS
1
:
TIME
SENS2
:
numeric_value
DEF
Space
SWE
?
1
2
3
4
Item Description/Default Range of Values
numeric_value The length of the time gated period in seconds.
– DEF: the default value is 100 µs
Units are resolved to 1 ns.
0 to 1 second
DEF
SENS2:SWE3:TIME 0.001 This command sets the length to 0.001
seconds.