MEASurement Commands 2
Keysight N8262A Programming Guide 133
CONFigure[1]|2|3|4[:SCALar][:POWer:AC]
[<expected_value>[,<resolution>[,<source list>]]]
This command is used on the specified window/measurement to set:
The expected power level of the measurement.
The resolution of the window/measurement.
The channel on which the measurement will be made.
Syntax
Parameters
Refer to “Optional Parameters” on page 126 for additional details on the
parameters in this command.
Space
expected_value
DEF
resolution
,
:POW
:AC
:SCAL
DEF
source list
,
CONF
1
2
3
4
Item Description/Default Range of Values
expected_value A numeric value for the expected power level. The units of
measurement are dBm and W. The default units are defined
by UNIT:POWer.
Sensor dependent.
DEF