10 SENSe Subsystem
358 Keysight N8262A Programming Guide
[SENSe[1]]|SENSe2:AVERage2:COUNt <numeric_value>
This command is used to enter the video filter length for the P-Series sensor.
Video filtering is applied to the traces. Successive traces are combined to reduce
noise without affecting the dynamic characteristic of the signal.
Syntax
Parameters
Example
Reset Condition
On reset, the filter length is set to 4.
:COUN
Space
?
numeric_value
DEF
SENS
1
AVER2
SENS2
:
Item Description/Default Range of Values
numeric_value A numeric value defining the filter length.
– DEF: the default value is 4.
1 to 256
[a]
DEF
[a] This is only implemented in powers of 2 (2
n
).
AVER2:COUN 16 This command enters a video filter length of
16 for Channel A.