9 PSTatistic Subsystem
334 Keysight N8262A Programming Guide
Error Messages
If no power sensor is connected, error –241 "Hardware missing" occurs.
If sensor/sensors connected are not P-Series sensors, error –241 "Hardware
missing" occurs.
If measurement speed setting is FAST, error –221 "Settings conflict" occurs.
If the acquisition mode is in continuous triggering or triggering source is set to
either INT1, INT2, or EXT, error –221 "Setting conflict" occurs.
If the parameter specified is less than 0.0 or more than 100.0, error –220
"Parameter error" occurs.
If no parameter is specified, error –109 "Missing parameter" occurs.