MEASurement Commands 2
Keysight N8262A Programming Guide 159
READ[1]|2|3|4[:SCALar][:POWer:AC]?
[<expected_value>[,<resolution>[,<source list>]]]
This command sets the specified window’s measurement function to single
channel with relative mode off, aborts then initiates the specified channel,
calculates the measurement result and places the result on the bus. The result is a
power based measurement and is expressed in the units defined by
UNIT[1]|2|3|4:POWer.
Syntax
Parameters
Refer to “Optional Parameters” on page 126 for additional details on the
parameters in this command.
INITiate:CONTinuous must be set to OFF, otherwise error –213, “INIT ignored”
occurs. If TRIGger:SOURce is set to BUS, error –214, “Trigger deadlock” occurs.
:POW
:AC
:SCAL
?
Space
expected_value
DEF
resolution
,
DEF
source list
,
READ
1
2
3
4