CALibration Subsystem 4
Keysight N8262A Programming Guide 235
Syntax
Parameters
Example
Reset Condition
On reset, the reference calibration factor is set to 100 %.
:RCF
Space
?
numeric_value
DEF
MIN
MAX
MIN
MAX
Space
CAL
1
2
Item Description/Default Range of Values
numeric_value A numeric value:
– DEF: the default is 100 %
– MIN: 1 %
– MAX:150 %
1.0 to 150.0 PCT
DEF
MIN
MAX
CAL1:RCF 98 This command enters a reference calibration
factor of 98 % to Channel A.