4 CALibration Subsystem
238 Keysight N8262A Programming Guide
Query
CALibration[1]|2:ZERO:AUTO?
The query always returns a value of 0.
Error Messages
If this command is set to ON and an 8480 Series or E-Series power sensor is
connected the error –241, “Hardware missing” occurs.
If zeroing was not carried out successfully the error –231, “Data Questionable;
ZERO ERROR” occurs. The error message will specify which channel failed
zeroing.