SENSe Subsystem 10
Keysight N8262A Programming Guide 351
Syntax
Example
Reset Condition
On reset, automatic averaging is enabled.
Query
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO?
The query enters a 1 or 0 into the output buffer indicating whether automatic filter
length is enabled or disabled.
1 is returned when automatic filter length is enabled
0 is returned when automatic filter length is disabled
NOTE
For most applications, automatic filter length selection
([SENSe[1]]|SENSe2:AVERage:COUNt:AUTO ON) is the best mode of operation.
However, manual filter length selection ([SENSe[1]]|SENSe2:AVERage:COUNt
<numeric_value>) is useful in applications requiring either high resolution or fast
settling times, where signal variations rather than measurement noise need
filtering, or when approximate results are needed quickly.
AVER:COUN:AUTO OFF This command disables automatic filter
length selection for Channel A.
SENS
1
:COUN
AVER 0|OFF
1|ON
Space
?
SENS2
:AUTO
:
ONCE