SENSe Subsystem 10
Keysight N8262A Programming Guide 371
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe <boolean>
This command is to enable and disable the use of the currently active sensor
calibration table (CSET1) or frequency dependent offset table (CSET2). When a
table has been selected and enabled, the calibration factors/offsets stored in it
can be used by specifying the required frequency using the
[SENSe[1]]|SENSe2:FREQuency command.
When the CSET1 command is set to ON, the reference calibration factor is taken
from the sensor calibration table and is used during calibration.
Syntax
Example
Reset Condition
On reset, the sensor calibration table and frequency dependent offset table are
not affected.
CORR:CSET1:STAT 1 This command enables the use of the
currently active sensor calibration table for
Channel A.
SENS
1
:CSET
CORR
SENS2
:
:STAT
0|OFF
1|ON
Space
?
1
:CSET2