EasyManuals Logo

Keysight Technologies P Series User Manual

Keysight Technologies P Series
673 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #386 background imageLoading...
Page #386 background image
10 SENSe Subsystem
386 Keysight N8262A Programming Guide
[SENSe[1]]|SENSe2:FREQuency[:CW|:FIXed] <numeric_value>
This command is used to enter a frequency. If the frequency does not correspond
directly to a frequency in the sensor calibration table, the power meter calculates
the calibration factor using linear interpolation. For 8480 Series Power Sensor the
power meter uses linear interpolation to calculate the calibration factor for the
frequency entered if [SENSe[1]]|SENSe2:CORRection:CSET:STATe is ON. For
P-Series and E-Series Power Sensor, the appropriate corrections are applied for
the frequency selected, dependant on the calibration data stored in the sensor’s
EEPROM.
Syntax
:CW
Space
?
numeric_value
DEF
MIN
MAX
MIN
MAX
Space
SENS
1
FREQ
SENS2
:
:FIX

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals