EasyManua.ls Logo

Keysight Technologies P Series

Keysight Technologies P Series
673 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SENSe Subsystem 10
Keysight N8262A Programming Guide 391
[SENSe[1]]|SENSe2:POWer:AC:RANGe <numeric_value>
This command is only valid when used with an E-Series Power Sensor. Its purpose
is to select one of two power ranges.
If 0 is selected, the power sensor’s lower range is selected
If 1 is selected, the power sensor’s upper range is selected
Setting a range with this command automatically switches
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO to OFF.
Syntax
Example
Reset Condition
On reset, the upper range is selected.
Query
[SENSe[1]]|SENSe2:POWer:AC:RANGe?
The query enters a 1 or 0 into the output buffer indicating the status of the power
sensor’s range.
1 is returned when the upper range is selected
0 is returned when the lower range is selected
POW:AC:RANG 0 This command sets the power sensor to it’s
lower range.
SENS
1
:AC
POW
SENS2
:
:RANG
Space
?
numeric_value

Table of Contents

Other manuals for Keysight Technologies P Series

Related product manuals